In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 16

Advertising
background image

Name

Command

Response

Start Concurrent

Start Concurrent CRC






Send Data





Additional Data

aC!

aCC!






aD0!





aD1! … aD9!

a004nn<CR><LF>

nn parameters will be available for reading by the Send
Data command within 4 seconds. No service request
will be sent when the parameters are ready. The
number of parameters returned is determined by the
SDI-12 configuration file in the same manner as a Start
Measurement command.

a<values><CR><LF> or
a<values><CRC><CR><LF>
The number and type of parameters returned is
determined by the SDI-12 configuration file in the same
manner as a Start Measurement command.

a<values><CR><LF> or
a<values><CRC><CR><LF>
The number and type of parameters returned is
determined by the SDI-12 configuration file in the same
manner as a Start Measurement command.

Additional Concurrent

Additional with CRC


Send Data


Additional Data

aC1! … aC9!

aCC1! … aCC9!


aD0!


aD1! … aD9!

a00000<CR><LF>

No additional concurrent measurements are started by
the device.

a<CR><LF> or
a<CRC><CR><LF>

No values are returned after an additional Concurrent
Measurement command.

Continuous Measurement

Continuous with CRC

aR0! … aR9!

aRC0! … aRC9!

a<CR><LF>

No values shall be returned after a Continuous
Measurement command.

16

Advertising