7.2 Parameter Registers

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

pH

If the parameter value is read when the sensor is not plugged in, it returns the sentinel value and a Data

Quality ID of 7 (sensor communication error).

If the parameter value calculates to a value outside of the range 0 to 14 pH, it shall return the sentinel

value and a Data Quality ID of 3 (error).

1538

2

R1

float

Measured value

1540

1

R1

ushort Parameter ID = 17 (pH)

1541

1

R1/W2

ushort Units ID

145 = pH (default)

1542

1

R1

ushort Data Quality ID

1543

2

R1/W3

float

Off line sentinel value (default = 0.0)

1545

1

R1

16 bits Available Units = 0x0001 (1)

pH/mV

If the parameter value is read when the sensor is not plugged in, it shall return the sentinel value and a

Data Quality ID of 7 (sensor communication error).

1546

2

R1

float

Measured value

1548

1

R1

ushort Parameter ID = 18 (pH mV)

1549

1

R1/W2

ushort Units ID

162 = mV (default)

1550

1

R1

ushort Data Quality ID

1551

2

R1/W3

float

Off line sentinel value (default = 0.0)

1553

1

R1

16 bits Available Units = 0x0002 (2)

ORP/mV

If the parameter value is read when the sensor is not plugged in, it shall return the sentinel value and a

Data Quality ID of 7 (sensor communication error).

1554

2

R1

float

Measured value

4556

1

R1

ushort Parameter ID = 19 (ORP)

51

Advertising