2 orp measured value, 4 calibration procedure—ph, 5 calibration procedure—orp – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 54: 1 sensor command timeouts, Orp measured value, Calibration procedure—ph, Calibration procedure—orp, Sensor command timeouts

Advertising
background image

7.3.2 ORP Measured Value

The measured value is derived as follows:

ORP = m

orp

× mV + b

orp

Where:

mV is the measured sensor millivolts

m

orp

and b

orp

are slope and offset respectively

ORP is the calculated value

7.4 Calibration Procedure—pH

1. Write the Calibration Mode On command (0xE000) to the Sensor Command

register.

2. Place the pH sensor into one or more pH standards.
3. Read the pH mV and temperature parameters at each pH standard.
4. Calculate new pH calibration values and write these values to their

corresponding registers.

5. Write the Calibration Update command (0xE001) to the Sensor Command

register. The sensor sets the Last User Calibration Date to the current date and
sets the Next User Calibration Date to zero (none required).

6. Optionally, read the Last User Calibration Time register, add the next calibration

interval, and write the result to the Next User Calibration Time register.

7. Write the Calibration Off command (0xE002) to the Sensor Command register to

place the sensor in normal operation.

7.5 Calibration Procedure—ORP

1. Write the Calibration Mode On command (0xE000) to the Sensor Command

register.

2. Place the ORP sensor into a standard.
3. Read the ORP mV and temperature parameters of the standard. Reading the

ORP parameter in the calibration will return raw ORP millivolts, without the
current calibration applied.

4. Calculate new ORP calibration values and write these values to their

corresponding registers.

5. Write the Calibration Update command (0xE001) to the Sensor Command

register. The sensor sets the last user calibration date to the current date and
sets the Next User Calibration Date to zero (none required).

6. Optionally, read the Last User Calibration Time register, add the next calibration

interval, and write the result to the Next User Calibration Time register.

7. Write the Calibration Off command (0xE002) to the Sensor Command register to

place the sensor in normal operation.

7.5.1 Sensor Command Timeouts

Both the Sensor Cal Defaults command (0xE003) and the Restore Sensor Defaults command
(0xE004) require a response timeout of 1000 milliseconds.

54

Advertising
This manual is related to the following products: