3 calibration registers—level, Calibration registers—level – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 46

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

Level

1054

2

R1

float

Measured value, L

m

1056

1

R1/W2

ushort Parameter ID

3 = level, depth (default)

4 = level, top of casing

5 = level, elevation

1057

1

R1/W2

ushort Units ID

33 = millimeters

34 = centimeters

35 = meters

37 = inches

38 = feet (default)

1058

1

R1

ushort Data Quality ID

1059

2

R1/W3

float

Off line sentinel value (default = 0.0)

1061

1

R1

16 bits Available Units = 0x0037 (55)

6.3 Calibration Registers—Level

Values in the calibration registers determine how sensor parameters are calculated.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

1100

2

R1/W3

float

Specific Gravity (default = 1.0)

1102

2

R1/W3

float

Pressure Offset, P

O

(default = 0.0)

1104

2

R1/W3

float

Level Reference, L

R

(default = 0.0)

1106

2

R1/W3

float

Pressure Reference, P

R

(default = 0.0)

46

Advertising
This manual is related to the following products: