14 probe register map layout, Probe register map layout – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 68

Advertising
background image

Code

Name

Description

0x94

Sensor Missing

Attempting a sensor operation on a sensor port

with no sensor connected.

0x95

Sensor Invalid

Attempting a sensor operation on a sensor port that has a

sensor that is not compatible with the sensor that

was originally configured.

0x96

Sensor Firmware

Attempting a sensor operation on a sensor

with no application code.

0xA0

Data Log Register

Attempting to write a register that is read-only during logging.

0xA1

Data Log Memory

Data log memory is full.

0xA2

Data Log Directory

Data log directory is full.

0xA3

Data Log Edit

Log configuration cannot be edited.

0xA4

Data Log Sequence

Invalid data log command sequence.

8.14 Probe Register Map Layout

Each probe and sensor will follow a common layout pattern which provides consistency and
improves code reuse between platforms. All Instrument registers are Holding Registers. The
common probe register map layout is as follows:

Probe Common

Registers

9000 - 9209

Sensor Common

Registers

1 - 38

Parameter 1

Registers

38 - 45

Parameter 2

Registers

46 - 53

Parameter N

Registers

xx – xx+8

Other Registers

xxxx-xxxx