In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 19

Advertising
background image

Name

Command

Response

Set Specific
Conductivity Alpha

aXCApd.d!

aOK<CR><LF> or
aERROR<CR><LF>

pd.d = alpha
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

Read Specific
Conductivity Alpha

aXCA!

apd.d<CR><LF>
pd.d = Alpha

Set Specific Conductivity
Reference Temperature

aXCTpd.d!

aOK<CR><LF> or
aERROR<CR><LF>
pd.d = reference temperature
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

Read Specific Conductivity
Reference Temperature

aXCT!

apd.d<CR><LF>
pd.d = reference temperature

Set TDS Units

aXDUnnn!

aOK<CR><LF> or
aERROR<CR><LF>

nnn = the TDS units ID
Three digits are required.

Read TDS Units

aXDU!

annn<CR><LF>
nnn = TDS units ID

Set TDS Factor

aXDFpd.d!

aOK<CR><LF> or
aERROR<CR><LF>

pd.d = conversion factor
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

Read TDS Factor

aXDF!

apd.d<CR><LF>
pd.d = conversion factor

Restore Conductivity
Factory Calibration




Send Data

aXCFC!





aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. Restores conductivity
sensor cell constant to its factory default. No
change to other parameter settings.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

19

Advertising
This manual is related to the following products: