8.29 Device Status

The device status register holds general status information. Each set bit represents a status
value. There are a limited number of standardized predefined status values that all devices will
support. These predefined status values are contained in the lower register. The upper register
is reserved for device specific status values.

Bit

Category

Description

0

Alarm

Sensor high alarm

1

Warning

Sensor high warning

2

Warning

Sensor low warning

3

Alarm

Sensor low alarm

4

Warning

Sensor calibration warning

5

Alarm

Sensor malfunction

6-7

N/A

Reserved

8

Status

Power management disabled

9

Status

Device off line

10

Alarm

Device hardware reset occurred

11

Alarm

Device malfunction

12

Status

No external power.

13

Warning

Low battery – battery capacity < 5%

14

Warning

Low memory – data log memory capacity < 5%

15

N/A

Reserved

16-31

N/A

Available for device-specific status

• Bits 0-7 of the device status register are reserved for sensor status. These bits are

the logical OR of bits 0-7 of the sensor status register in each sensor connection.

• Bits 8-15 of the device status register are reserved for common device status. Any bit

in this range that is not applicable to a device will be set to zero.

• Bits 16-31 of the device status register are available for device-specific status. Any

bit in this range that is not utilized by a device will be set to zero.