1 ph measured value, Ph measured value – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 53

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

1641

2

R1/W3

float

Slope, m

2

in pH/mV

(default = -0.016903313)

1643

2

R1/W3

float

Offset, b

2

in pH (default = 7.0)

1645

1

R1/W3

ushort ORP solution type (default = 0)

1646

2

R1/W3

float

mV value of solution (default = 0)

1648

2

R1/W3

float

mV response in solution (default = 0)

1650

2

R1/W3

float

Temperature of solution, T

orp

in °C

(default = 25)

1652

2

R1/W3

float

Slope, m

orp

in mV/mV (default = 1.0)

1654

2

R1/W3

float

Offset, b

orp

in mV

(default = 0.0, limit = +/-1000)

7.3.1 pH Measured Value

The measured value is derived from either one line segment (created from a one-point or two-
point calibration) or two line segments (created from a three-point calibration).

For one line segment:

pH = (T

1K

/ T

K

) × m1 × mV + b

1

For two line segments:

mV > mV2: pH = (T

1K

/ T

K

) × m

1

× mV + b

1

mV <= mV

2

: pH = (T

2K

/ T

K

) × m

2

× mV + b

2

Where:

mV is the measured sensor millivolts

T

K

is the measured temperature in °K

T

1K

is the calibration temperature converted to °K. (T

1K

= T

1

+ 273.15)

T

2K

is the calibration temperature converted to °K. (T

2K

= T

2

+ 273.15)

m

1

and b

1

are slope and offset for segment one respectively

m

2

and b

2

are slope and offset for segment two respectively

pH is the calculated result