8. Optionally, read the Last User Calibration Time, add the next calibration interval,

and write the result to the Next User Calibration Time register.

9. Write the Calibration Off command (0xE002) to the Sensor Command register to

place the sensor in normal operation.

7. Modbus Registers – pH/ORP Sensor

The Sensor Map Data Register Offset points to the first register in the sensor data header block.
The current value for this sensor data register map version is 1.

The pH/ORP sensor registers remain valid even if the sensor is not installed.

7.1 Header Registers

The sensor map data register offset points to the first register in the sensor data header block.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

1501

1

R1/W4

ushort Sensor ID

0 = not installed

27 = pH/ORP

1502

2

R1/W4

ulong

Sensor serial number (0 – 65535)

1504

1

R1

16 bits Sensor status

1505

3

R1/W4

time

Last factory calibration

1508

3

R1

time

Next factory calibration

(0 = none required)

1511

3

R1

time

Last user calibration

1514

3

R1/W2

time

Next user calibration

(0 = none required)

1517

1

R1

ushort Warm-up time =

1300 milliseconds

1518

1

R1

ushort Fast sample rate =

1300 milliseconds

1519

1

R1

ushort Number of sensor parameters

(N = 3)

1520

1

R1/W3

ushort Alarm/warning parameter number

(1 – N, default = 1)

49