In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 40

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

0567

2

R1/W3

float

Off line sentinel value

(default = 0.0)

0569

1

R1

16 bits

Available Units = 0x0001 (1)

Total Dissolved Solids

The default units for TDS are ppt. Conversion to other units is as follows.

ppm = ppt * 1000

TDS is calculated from specific conductivity using the following equation.

TDS = CF

TDS

* SC

Where CF

TDS

is the TDS conversion factor in ppm units as specified in the sensor calibration registers.

The default conversion factor is 0.65.

0570

2

R1

float

Measured value, TDS

0572

1

R1

ushort

Parameter ID = 13 (TDS)

0573

1

R1/W2

ushort

Units ID

113 = parts per million

114 = parts per thousand

(default)

0574

1

R1

ushort

Data Quality ID

0575

2

R1/W3

float

Off line sentinel value

(default = 0.0)

0577

1

R1

16 bits

Available Units = 0x0003 (3)

Resistivity

Resistivity is calculated from actual conductivity using the following equation.

R (ohm-cm) = 10

6

/ AC (uS/cm)

If the actual conductivity is zero, R is reported as 10,000,000 ohm-cm.

0578

2

R1

float

Measured value, R

0580

1

R1

ushort

Parameter ID = 11(resistivity)

0581

1

R1/W2

ushort

Units ID 81 = ohm-cm (default)

0582

1

R1

ushort

Data Quality ID

40