2 sensor serial number, 3 sensor status, 4 last factory calibration – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 29: Sensor serial number, Sensor status, Last factory calibration

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

0020

1

R1/W3

ushort Alarm/warning parameter number

(1 – N, default = 1)

0021

1

R1/W3

16 bits Alarm and warning enable bits

(default = 0)

Bit 0 = High alarm enabled

Bit 1 = High warning enabled

Bit 2 = Low warning enabled

Bit 3 = Low alarm enabled

Bit 4 = Sensor calibration warning

0022

2

R1/W3

float

High alarm set value (default = 0.0)

0024

2

R1/W3

float

High alarm clear value (default = 0.0)

0026

2

R1/W3

float

High warning set value (default = 0.0)

0028

2

R1/W3

float

High warning clear value

(default = 0.0)

0030

2

R1/W3

float

Low warning clear value

(default = 0.0)

0032

2

R1/W3

float

Low warning set value (default = 0.0)

0034

2

R1/W3

float

Low alarm clear value (default = 0.0)

0036

2

R1/W3

float

Low alarm set value (default = 0.0)

4.2 Sensor Serial Number

This field returns the serial number of the installed sensor cap. The probe returns zero if the
sensor cap is not installed or is expired.

4.3 Sensor Status

Bit 9 of the sensor status register is set to indicate when the sensor cap is not installed.

4.4 Last Factory Calibration

This field returns the start time of the installed sensor cap. If the cap has not been used, the cap
manufactured time is returned. The probe returns zero if the sensor cap is not installed.

29

Advertising