2 parameter registers—conductivity, Parameter registers—conductivity – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 38

Advertising
background image

5.2 Parameter Registers—Conductivity

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

Actual Conductivity

0538

2

R1

float

Measured value, AC

0540

1

R1

ushort

Parameter ID = 9

(actual conductivity)

0541

1

R1/W2

ushort

Units ID

65 = microsiemens

per centimeter (default)

66 = millisiemens

per centimeter

0542

1

R1

ushort

Data Quality ID

0543

2

R1/W3

float

Off line sentinel value

(default = 0.0)

0545

1

R1

16 bits

Available Units = 0x0003 (3)

Temperature

0546

2

R1

float

Measured value

0548

1

R1

ushort

Parameter ID = 1

(temperature)

0549

1

R1/W2

ushort

Units ID

1 =

°C (default)

2 =

°F

0550

1

R1

ushort

Data Quality ID

0551

2

R1/W3

float

Off line sentinel value

(default = 0.0)

0553

1

R1

16 bits

Available Units = 0x0003 (3)