In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 39

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

Specific Conductivity

The default units for specific conductivity are uS/cm.

Conversion to other units is as follows.

mS/cm = uS/cm / 1000

Specific conductivity is calculated from actual conductivity and temperature using the following

equation.

SC = AC * (

β

0

+

β

1

T +

β

2

T

2

+ … +

β

7

T

7

) / (1 +

α (T-Tref))

Where T

ref

,

α, and β

0-7

are specified in the sensor calibration registers. The factory default coefficients

calculate specific conductivity per Standard Methods 2510B.

0554

2

R1

float

Measured value, SC

0556

1

R1

ushort

Parameter ID = 10

(specific conductivity)

0557

1

R1/W2

ushort

Units ID

65 = microsiemens

per centimeter (default)

66 = millisiemens

per centimeter

0558

1

R1

ushort

Data Quality ID

0559

2

R1/W3

float

Off line sentinel value

(default = 0.0)

0561

1

R1

16 bits

Available Units = 0x0003 (3)

Salinity

Salinity is calculated using actual conductivity and temperature.

0562

2

R1

float

Measured value, S

0564

1

R1

ushort

Parameter ID = 12 (salinity)

0565

1

R1/W2

ushort

Units ID

97 = Practical Salinity Units

PSU (default)

0566

1

R1

ushort

Data Quality ID

39