5 result variables, Result varia- bles – Metrohm tiBase 1.1 User Manual

Page 36

Advertising
background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

tiBase 1.1

Variable
name

(PC-Control)

Variable
name

( tiBase 1.1)

Description

Command

.MSP

.SVM

Start volume measured value
(volume which was added
according to the start measured
value given in the start condi-
tions) in mL

DET, MET

.MSS

.SVS

Start volume measured value
(volume that was added accord-
ing to the start condition "start
slope") in mL

DET, MET

.MSV

.SVT

Total start volume (volume that
was added in dependence on all
three start conditions) in mL

DET, MET, SET, KFT, STAT

.TITER

.TITER

Titer value of the solution used
for the command (number)

DET, MET, SET, KFT, STAT,
STDADD dos, STDADD
auto, ADD, DOS, LQH

.MVA

.VAR

Variance of the result calculated
from the standard addition curve

STDADD

.MCV

.VOL

Dosed volume

STDADD, ADD, LQH

1) HNP = Half neutralization potential

2.4.3.5

Result variables

Dialog window: Formula Editor

Variables

Result variables are method-specific and are defined in the CALC com-
mand of the method. The Variables field of the Formula editor lists all
Result variables which are available for the current method.

Syntax
'RS.Result name.Variable identification'

Examples: 'RS.RS01.VAL' (='RS.RS01'), 'RS.testTime.UNI'

You can select the method variables directly in the formula editor under
Variables

Result Variables in order to avoid syntax errors.

Result variables

Variable
name

(PC-Control)

Variable
name

( tiBase 1.1)

Description

Command

.R1 to .R9

[.VAL]

Result value ( .VAL optional;
number)

CALC

SSA#

.ASD

Absolute standard deviation for
the result (number)

CALC

Advertising