3 Configuration

tiBase 1.1

■■■■■■■■

87

Selection

on | off

Default value

off
Recommended setting: off

Automatic logout after

If this check box is activated, then the user will be logged out automati-
cally when no operating functions have been carried out with the key-
board or the mouse within this time. After this automatic logout, only a
member of the same user group as the one to which the previously log-
ged-off user belongs or a user with administrator rights can log in again.

Selection

on | off

Default value

off
Recommended setting: off

Input range

1 - 60 min

Default value

10 min
Recommended setting: 10

Password policies

Note

If the Password monitoring by Windows option is switched on,
then the parameters for password protection are deactivated. Compli-
ance with the recommended settings must then be ensured by the
administrator in Windows.

Enforce unique password

Activating this check box ensures that a user can use a given password
only once.

Selection

on | off

Default value

off
Recommended setting: on (can no longer be edi-
ted)

Enforce use of special characters

Activating this check box ensures that the password must contain at
least one special character (@, #, ~ , etc.).

Selection

on | off

Default value

off
Recommended setting: off