1 introduction, 1 instrument description, Instrument description – Metrohm 861 Advanced Compact IC User Manual

Page 11

Advertising
background image

1.1 Instrument description

861 Advanced Compact IC / Instructions for Use 8.861.1033

1

1 Introduction

1.1

Instrument

description

The 861 Advanced Compact IC is a PC-controlled system for ion

chromatographic analyses. The three following versions are available:

2.861.0010 without suppressor

2.861.0020 with suppressor module «MSM II»

2.861.0040 with suppressor module «MSM II»

and 853 CO

2

Suppressor (MCS)

The extremely compact housing of the 861 Advanced Compact IC con-

tains everything needed to carry out ion chromatography at the highest

quality level:

Injection valve – for individual injections or for use with a sample

changer such as the Metrohm 838 Advanced Sample Processor

High-pressure pump – extremely low-pulsation double piston

pump with a flow range from 0.2

2.5 mL/min and a maximum

pressure of 35 MPa (350 bar)

Pulsation dampener – even with low-level pressure variations the

pulsation dampener reliably protects the column against damage

Column chamber – the perfect insulation of the housing provides

not only thermally stable conditions for the separation column but

also shields the system against electromagnetic interference

Columns – whether anion columns with or without suppression,

separation columns for cations or organic acids – the 861 Ad-

vanced Compact IC can accommodate them all

Suppressor module – the Metrohm suppressor module (MSM) is

already integrated in the versions 2.861.0020 and 2.861.0040 –

pressure-resistant, with fully automatic regeneration, highest per-

formance and optimal reproducibility

Peristaltic pump – integrated two-channel peristaltic pump with a

flow rate of 0.4

0.5 mL/min for regeneration and rinsing of the

suppressor module «MSM II» built into the versions 2.861.0020 and

2.861.0040.

Detector – conductivity detector with outstanding temperature

stability. The detector temperature varies by less than 0.01°C and

can be optimally adapted to the ambient conditions.

853 CO

2

Suppressor – in Version 2.861.0040 the 853 CO

2

Sup-

pressor (MCS) is built in downline from the suppressor module

«MSM II». The·853 CO

2

Suppressor (MCS) removes the CO

2

from

the eluent. This reduces the background conductivity, improves de-

tection sensitivity and minimizes the injection and carbonate peak.

Advertising