2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

32

Anion columns

100

µ

L

Cation columns

10

µ

L

Columns for suppressor technique 20

µ

L

The following sample loops are installed in the 861 Advanced Compact
IC:

Version

Sample loop

Volume

2.861.0010 6.1825.210

(PEEK) 20

µ

L

2.861.0020 6.1825.210

(PEEK) 20

µ

L

2.861.0040 6.1825.210

(PEEK) 20

µ

L

If desired, the built-in sample loop can be replaced by one of the sam-
ple loops available as an option (see Section 5.3.1).

2.7.4

Installation of the separating column without suppressor
module «MSM II»

With the 2.861.0010 Compact IC without suppressor module «MSM II»,
the IC separating column is installed as follows (see Figure 14):

1 Connect column to injection valve

Remove end caps from column 73.

without precolumn:

Screw inlet end of separating column 73 (note flow direction)

to column connection capillary 26 mounted on the injection

valve.

with precolumn:

See relevant leaflet.

2 Rinse column

Place a beaker beneath the column outlet.

Open software window for manual system control.

If necessary, modify

Flow rate

to the value suited for the inser-

ted separating column and click on

<Send to unit>

to send this

value to the 861 Advanced Compact IC.

Switch on high-pressure pump (

IC pump

) by clicking

<On>

and rinse column with eluent for approx. 10 min.

Switch off high-pressure pump by clicking

<Off>

.

3 Connect column to detector block

Screw outlet end of separating column 73 to the inlet capillary

44 permanently mounted on the detector block 45.

4 Column attachment

Insert one or two column holders 74 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 25 and fasten separating

column 73 in the column holder 74.