2.8 Suppressor module «MSM II»

861 Advanced Compact IC / Instructions for Use 8.861.1033

39

2) from above and pull it sufficiently far into the interior.

Mount a coupling 77 (6.2744.030) to the inlet end of the rear

pump tubing 78.

Mount a 6.2744.010 Compression fitting at the end of the as-

pirating tubing 76 and screw this compression fitting on to

the coupling 77 (see Figure 16).

3 Prepare supply bottle for H

2

O

Prepare rinsing solution suited for the desired application and

separating column (normally dist. H

2

O).

Fill rinsing solution into supply bottle 89 (amber glass, 1 L)

and label the bottle.

Screw bottle attachment 88 on to supply bottle 89.

Place supply bottle 89 beside the other supply bottle on the

bottle rack 6 on the 861 Advanced Compact IC (see Figure

1).

4 Connect aspirating tubing for H

2

O

Prepare aspirating tubing 75: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally approx. 120

cm).

Pull one end of aspirating tubing 75 through a PEEK com-

pression fitting 54 (6.2744.010) so that approx. 30 cm of the

tubing projects.

Screw PEEK compression fitting 54 with tubing into one

opening of bottle attachment 88 attached to the rinsing solu-

tion storage bottle and tighten it so that the tubing is firmly

held.

Insert the free end of aspirating tubing 75 into one of the

openings in 8 of the 861 Advanced Compact IC (see Figure

2) from above and pull it sufficiently far into the interior.

Mount a coupling 77 (6.2744.030) to the inlet end of the front

pump tubing 79.

Mount a 6.2744.010 Compression fitting at the end of the as-

pirating tubing 75 and screw this compression fitting on to

the coupling 77 (see Figure 16).

Advertising