2.6 High-pressure pump

861 Advanced Compact IC / Instructions for Use 8.861.1033

29

Select the

Connect to workplace

item of the

Control

menu in this

window.

3 Set flow rate to 2 mL/min

Double-click the 861 icon in the system window to open the

window for manual control of the 861 Advanced Compact IC

(see below).

Set the flow rate to

2 mL/min

in the

Flow

field.

Click to

<Send to unit>

to send this value to the 861 Advanced

Compact IC.

4 Deaerate pump

Make sure that the aspirating tubing 63 for the high-pressure

pump has been immersed into the eluent.

Click the

<On>

button for

IC pump

to switch on the high-pres-

sure pump.

Use the syringe inserted into connection 2 to aspirate air until

eluent flows bubble-free into the syringe.

Click the

<Off>

button for

IC pump

to switch off the high-pres-