Metrohm 861 Advanced Compact IC User Manual
Page 39
2.6 High-pressure pump
861 Advanced Compact IC / Instructions for Use 8.861.1033
29
•
Select the
Connect to workplace
item of the
Control
menu in this
window.
3 Set flow rate to 2 mL/min
•
Double-click the 861 icon in the system window to open the
window for manual control of the 861 Advanced Compact IC
(see below).
•
Set the flow rate to
2 mL/min
in the
Flow
field.
•
Click to
<Send to unit>
to send this value to the 861 Advanced
Compact IC.
4 Deaerate pump
•
Make sure that the aspirating tubing 63 for the high-pressure
pump has been immersed into the eluent.
•
Click the
<On>
button for
IC pump
to switch on the high-pres-
sure pump.
•
Use the syringe inserted into connection 2 to aspirate air until
eluent flows bubble-free into the syringe.
•
Click the
<Off>
button for
IC pump
to switch off the high-pres-