3 Operation

861 Advanced Compact IC / Instructions for Use 8.861.1033

64

Outputs

The

Outputs

tab of the hardware settings window defines remote output

signals to be set automatically if specific events occur.

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "INJECT"

position.

Fill

Switching of the injection valve to the "FILL" po-

sition.

Suppressor step

Switching of the suppressor module to the next

position.

Pump stop (P max/min)

Pump stopped because pressure limits are ex-

ceeded.

Hardware error

Hardware error detected at the 861 Advanced

Compact IC (high-pressure pump, injection

valve, or suppressor not working correctly).

Alarm leak detector

Internal

leak.

Remote line

Set remote output lines 1

8.

Selection:

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(pulse output)

*

(leave line in current status)

PowerOn values

Display of the power on startup values for
remote output lines set on the

Hardware

tab.

Pulse length

Length of a pulse in ms.