Metrohm 861 Advanced Compact IC User Manual

Page 85

Advertising
background image

4.2 Maintenance and servicing

861 Advanced Compact IC / Instructions for Use 8.861.1033

75

the upper opening in the valve housing. The individual com-

ponents are rinsed with dist. water and/or acetone, and the

sapphire sphere cleaned with a paper towel. The valve is then

reassembled in accordance with Figure 21. The components

of the inlet and outlet valves are identical, they are distin-

guished only by the positioning of the sapphire sleeve 107

and the ceramic holder 109 (see Figure 21).

Valves that fail to function faultlessly after such cleaning must

be replaced.

In the reinstallation of the inlet valve 99 or the outlet valve 100

on no account must the two outwardly identical valves be in-

terchanged. To determine which valve is which, note that the

liquid flows through the pump head from the bottom up. The

flow direction of the valves can be checked simply by blowing

through the clean valve. Both valves are installed with the

black face in the direction of the pump head (see Figure 19).

If by mistake an inlet valve

99

is installed instead of the outlet valve

100

, an extreme pressure buildup occurs within the working cylinder,

which is not detected by the pressure transducer and will destroy the

piston seal

98

!

6 Mounting the pump head

Reassemble the components of the pump head 40 as shown

in Figure 19. Tighten the screw 90 by hand. First screw in pis-

ton cartridge 94 manually until the stop is reached and then

use a wrench to turn it through a furt

her 15°. Firmly tighten the

two valve screw holders 101 with a wrench.

Reattach pump head 40 to the pump with the help of the 4

fixing screws. Firmly tighten them with the 6.2621.030 Allen

key.

To ensure that the pump head is not positioned wrongly, the

holes at the rear for the clamping bolts have different depths,

i.e. 1 clamping bolt is longer than the rest. The deepest hole

must naturally accommodate the longest bolt. If this is not the

case, the pump will not function properly.

Screw connection capillary 39 to pump head 40 (see

Figure 3 / Figure 4 / Figure 5).

Connect aspirating tubing 63 to aspirating capillary 38 at

the pump head 40.