2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

38

50 Snap-action lever

for releasing the tubing cartridge

51 Pump drive

roller head with contact rollers

52 Mounting pin

for attaching the tubing cartridges

73 Separating column

74 Column holder (6.2027.0X0)

75 Aspirating tubing for H

2

O

76 Aspirating tubing for H

2

SO

4

77 Coupling (6.2744.034)

78 Pump tubing (6.1826.110) for

H

2

SO

4

79 Pump tubing (6.1826.110) for H

2

O

81 PEEK coupling with filter and tub-

ing security device 6.2744.180

82 Suppressor inlet capillary for

eluent

83 Suppressor outlet capillary for elu-

ent

84 Suppressor inlet capillary for H

2

O

85 Suppressor inlet capillary for

H

2

SO

4

86 Suppressor outlet capillary for H

2

O

87 Suppressor outlet capillary for

H

2

SO

4

2.8.3

Connection of supply bottles

The supply lines for the regeneration and rinsing solution between the
storage bottles and the peristaltic pump are installed as follows (see
Figure 17):

1 Prepare supply bottle for H

2

SO

4

Prepare regeneration solution suited for the desired applica-

tion and separating column (normally 50 mmol/L H

2

SO

4

).

Fill regeneration solution into supply bottle 89 (amber glass,

1 L) and label the bottle.

Screw bottle attachment 88 on to supply bottle 89.

Place supply bottle 89 at one of the rear positions of the bot-

tle rack 6 on the 861 Advanced Compact IC (see Figure 1).

2 Connect aspirating tubing for H

2

SO

4

Prepare aspirating tubing 76: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally approx. 120

cm).

Pull one end of aspirating tubing 76 through a PEEK com-

pression fitting 54 (6.2744.010) so that approx. 30 cm of the

tubing projects.

Screw PEEK compression fitting 54 with tubing into one

opening of bottle attachment 88 attached to the regeneration

solution storage bottle and tighten it so that the tubing is

firmly held.

Insert the free end of aspirating tubing 76 into one of the

openings in 8 of the 861 Advanced Compact IC (see Figure