6 connections, 2 maintenance and servicing, 1 general information – Metrohm 861 Advanced Compact IC User Manual

Page 79: Connections, Maintenance and servicing, General information

Advertising
background image

4.2 Maintenance and servicing

861 Advanced Compact IC / Instructions for Use 8.861.1033

69

Operation
The suppressor module «MSM II» comprises a total of 3 suppressor

units which are in turn used for suppression, regenerated with sulfuric

acid and rinsed with water. To record every new chromatogram under

comparable conditions, work is normally carried out with freshly regen-

erated suppressor. Switching takes place automatically together with

the valve switching.

Suppressor units must never be regenerated with H

2

SO

4

in the same

flow direction in which the eluent was pumped. This is why you should

always connect the inlet and outlet capillaries according to the ar-

rangement shown in Figure 18 in Section 2.8.4.

Suppressor module «MSM II» must never be switched to the next posi-

tion in a dry condition as this could cause a blockage. Each time be-

fore the suppressor module «MSM II» is switched on one further step

the three suppressor units must be rinsed for at least 30 min with elu-

ent, regenerating and rinsing solutions.

Maintenance
In the event of reduced capacity or high counter-pressure the suppres-

sor module «MSM II» must be regenerated (Section 4.2.7), cleaned

(Section 4.2.8) or exchanged (Section 4.2.9).

4.1.6 Connections

All connections between injection valve, column and detector must be

as short as possible, have a low dead volume and be absolutely tight.

The PEEK capillary after the detector block must be free from constric-

tion (the measuring cell is tested to 5 MPa = 50 bar back pressure).

Use PEEK capillaries with 0.25 mm i.d. for the capillary connection be-

tween injection valve 28 and detector block 45.

4.2 Maintenance

and

servicing

4.2.1 General

information

Care
The 861 Advanced Compact IC requires proper care and attention.

Excessive contamination of the instrument could possibly lead to mal-

functions and a shorter service life of the inherently rugged mechanical

and electronic parts.
For protection against escaping liquids the two drain tubes for the inner

compartment (Section 2.3.3) and for the bottle rack (Section 2.3.4) must

be mounted.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Advertising