4 peristaltic pump, 5 suppressor module «msm ii, Peristaltic pump – Metrohm 861 Advanced Compact IC User Manual

Page 78: Suppressor module «msm ii

Advertising
background image

4 Notes - Maintenance - Faults

861 Advanced Compact IC / Instructions for Use 8.861.1033

68

low buffering capacity one should preferably use a CO

2

absorber (see

Section 2.6.3).
The supply vessel containing the eluent must be closed as tightly as

possible to avoid excessive evaporation. This is primarily important with

eluents containing organic solvents (e.g. acetone), the evaporation of

which can lead to drifts in the long term. If work is performed in a very

sensitive range, even if one drop of condensate falls back in the eluent

this can cause a noticeable change in the background conductivity.

Influence of various parameters on anion columns

Concentration:

An increase in the concentration usually leads

to shorter retention times and quicker separa-

tion, but also to a higher background conduc-

tivity.

pH:

pH alterations lead to shifts in the dissociation

equilibrium and thus to changes in the reten-

tion times.

Organic modifiers:

Addition of an organic solvent (e.g. methanol,

acetone, acetonitrile) to aqueous eluents

generally accelerates lipophilic ions.

Eluent change
When changing the eluent you must make sure that no precipitation

can occur. This means that solutions used in sequence must be misci-

ble. If the system needs to be rinsed with organic solvents then it may

be necessary to use several solvents with increasing or decreasing

lipophilicity (e.g. water

acetone

chloroform).

4.1.4 Peristaltic

pump

The pump tubings used by the peristaltic pump are consumable mate-

rial with a limited lifetime and should be exchanged at regular intervals

(approx. every 4 weeks under continuous use; see Section 4.2.10).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in Section 2.10.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges 47 should

be raised completely by loosening the snap-action lever 50 on the right-

hand side (the set contact pressure remains unchanged).

4.1.5

Suppressor module «MSM II»

Protection
To avoid contamination of the suppressor module «MSM II» by foreign

particles or bacterial growth, a filter 6.2821.130 58 (part of the "PEEK

Coupling 81 with filter and tubing security device" 6.2744.180 ) must be

installed between peristaltic pump and inlet capillaries of the suppres-

sor module «MSM II» (see Section 2.8.2).

Advertising