2.11 Connection of external devices

861 Advanced Compact IC / Instructions for Use 8.861.1033

51

On Version 2.861.0040 the 853 CO

2

Suppressor is already

connected to Remote interface

21

. Connect the 6.2141.190

Cable to the rear of the Remote connector.

2 Tubing connection 813 – Injection valve

Screw off aspirating tubing 29 from injection valve 28 and

remove from feedthrough 4.

Cut the PEEK capillary tubing mounted on the 813 Compact

Autosampler to the required length (so that it can be connec-

ted to injection valve 28 of the 861).

Pull the free end of the PEEK capillary tubing through open-

ing 8 of the 861 Advanced Compact IC and use a 6.2744.010

PEEK Compression fitting to screw it onto injection valve 28.

3 Tubing connection Injection valve– Waste

Insert 6.2744.020 Coupling (from the 861 accessories) into

connection 3 of the 861 Advanced Compact IC.

Screw aspirating tubing 29 into the 6.2744.020 Coupling and

insert into the waste container.

4 Setting a method on the 813

Use the 813 keyboard on the 813 Compact Autosampler to

set the required method (see 813 Instructions for Use).

5 Creating a time program for the 861

In the main window of the «IC Net» software select

File / Open /

System

. Open the system for the 861 Advanced Compact IC

that you have created in Section 2.5.4.

Open the window with the system settings of the 861 (with a

double-click on the 861 icon or via the right-hand mouse key

open

).

Open the

Program

tab.

Create a time program:

The Remote command sends a signal to the 813 via Remo-

te line 3. The function of this signal depends on the method