2.11 Connection of external devices
861 Advanced Compact IC / Instructions for Use 8.861.1033
51
On Version 2.861.0040 the 853 CO
2
Suppressor is already
connected to Remote interface
. Connect the 6.2141.190
Cable to the rear of the Remote connector.
2 Tubing connection 813 – Injection valve
•
Screw off aspirating tubing 29 from injection valve 28 and
remove from feedthrough 4.
•
Cut the PEEK capillary tubing mounted on the 813 Compact
Autosampler to the required length (so that it can be connec-
ted to injection valve 28 of the 861).
•
Pull the free end of the PEEK capillary tubing through open-
ing 8 of the 861 Advanced Compact IC and use a 6.2744.010
PEEK Compression fitting to screw it onto injection valve 28.
3 Tubing connection Injection valve– Waste
•
Insert 6.2744.020 Coupling (from the 861 accessories) into
connection 3 of the 861 Advanced Compact IC.
•
Screw aspirating tubing 29 into the 6.2744.020 Coupling and
insert into the waste container.
4 Setting a method on the 813
•
Use the 813 keyboard on the 813 Compact Autosampler to
set the required method (see 813 Instructions for Use).
5 Creating a time program for the 861
•
In the main window of the «IC Net» software select
File / Open /
System
. Open the system for the 861 Advanced Compact IC
that you have created in Section 2.5.4.
•
Open the window with the system settings of the 861 (with a
double-click on the 861 icon or via the right-hand mouse key
open
).
•
Open the
Program
tab.
•
Create a time program:
The Remote command sends a signal to the 813 via Remo-
te line 3. The function of this signal depends on the method