2.3 Attaching the accessories

861 Advanced Compact IC / Instructions for Use 8.861.1033

19

These PEEK connectors can also be used to connect 6.1822.010 PTFE

microcapillaries (i.d. = 0.3 mm). Proceed as follows:

Capillaries fitted with new connectors must have a perfectly flat cut

surface. To cut PEEK or PTFE capillaries it is best to use the

6.2621.080 Capillary tubing cutter.

1 Mount compression fitting

Slide a compression fitting 54 (6.2744.010) or a compression

fitting 55 (6.2744.070) over the end of the capillary 56 to be

fastened as shown in Figure 9.

2 Insert capillary in connection

Push capillary end in the corresponding connection as far as

it will go (to avoid dead volume).

3 Tighten compression fitting

Tighten compression fitting 54 or 55 by hand (never use

tools).

Figure 9: Connectors for capillaries

2.3.6

Filter unit PEEK

One 6.2821.120 Filter unit PEEK (see Figure 10) is installed between

the high-pressure pump and injection valve 28. This filter unit serves to

avoid contamination by abrasive particles of the piston seals.
The Filter unit PEEK 34 consists of the filter-housing 59, the filter-screw

57 and the 6.2821.130 Filter 58. For the connection of capillaries 56

PEEK compression fittings 54/55 (6.2744.010 or 6.2744.070) must be

used. New filters 58 are available as an option with the ordering number

6.2821.130 (set of 10).

When connecting the filter unit please note the flow direction arrow

printed on the housing.

54 Compression fitting

(6.2744.010)

55 Compression fitting

(6.2744.070)

56 Capillary

6.1831.010 PEEK capillary or

6.1822.010 PTFE microcapillary

54

56

55