3.3 Settings

861 Advanced Compact IC / Instructions for Use 8.861.1033

57

Suppressor

Suppressor module «MSM II»

(inactive with version 2.861.0010)

<Step>

Switch the suppressor module to the next posi-

tion. The time since the last switching of the

suppressor module is displayed in the field be-

side the

<Step>

button.

IC pump

High-pressure

pump

<On>

Start high-pressure pump.

<Off>

Stop high-pressure pump.

Peristaltic pump

Peristaltic

pump

(inactive with version 2.861.0010)

<On>

Start peristaltic pump.

<Off>

Stop peristaltic pump.

CO

2

Suppressor

853

CO

2

Suppressor

(only displayed with version 2.861.0040)

<On>

Start 853 CO

2

Suppressor.

<Off>

Stop 853 CO

2

Suppressor.

System startup values

The system startup values are sent and applied

to the 861 Advanced Compact IC each time the

system is connected, a determination is started,

or the values are sent manually with

<Send to

unit>

.

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

35.0 MPa

If the

Flow

is higher than 1.5 mL/min, the defined

Pressure max

must not exceed 25.0 MPa (250 bar).

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

35.0

MPa

Full scale,

µ

S/cm

Startup value for full scale range.

Selection:

50

,

250

,

1000, 5000

µ

S/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1