2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

40

Figure 17: Connection of supply bottles

2.8.4

Connection of the suppressor module «MSM II»

The three inlets and outlets numbered 1...3 on the suppressor module
«MSM II» 46 each have 2 permanently mounted PTFE capillaries, which
must be connected as follows (see Figure 16 and Figure 18).

1 Inlet capillary for eluent

Screw inlet capillary 82 marked with "Eluent" of suppressor

module «MSM II» 46 to outlet end of separating column 73

using a 6.2744.010 Compression fitting.

2 Outlet capillary for eluent

Screw outlet capillary 83 marked with "Detector" of suppres-

sor module «MSM II» 46 to coupling 31 using a 6.2744.010

Compression fitting.

Screw inlet capillary 44 of detector block 45 to other end of

coupling 31.

For Version 2.861.0040 the outlet capillary

83

(marked with

"Detector") of suppressor module «MSM II»

46

should be con-

nected to the inlet

Eluent in

of the 853 CO

2

Suppressor (see

Section 2.9).

3 Inlet capillary for H

2

SO

4

Attach inlet capillary 85 marked with "H

2

SO

4

" of suppressor

module «MSM II» 46 using a 6.2744.010 Compression fitting

to the PEEK Coupling 81 connected to the rear pump tubing

78.

4 Outlet capillary for H

2

SO

4

Pull outlet capillary 87 marked with "Waste" of the suppressor

54 Compression fitting

(6.2744.010)

75 Aspirating tubing for H

2

O

76 Aspirating tubing for H

2

SO

4

88 Bottle attachment

(6.1602.150)

89 Supply bottle 6.1608.023

75/76

54
88

89