Figure 14, Installation of column without suppressor, Led as follows – Metrohm 861 Advanced Compact IC User Manual

Page 43: Figure 14 ), Column 73, 73 (note, 73 to the i, Insert one or two column holders 74 (6.20, Column 73 i

Advertising
background image

2.7 Precolumns and separating columns

861 Advanced Compact IC / Instructions for Use 8.861.1033

33

Figure 14: Installation of column without suppressor

2.7.5

Installation of the separating column with suppressor
module «MSM II»

With the versions 2.861.0020 and 2.861.0040 Compact IC with sup-

pressor module «MSM II», the IC separating column is first connected to
the injection valve or precolumn. The connection to the suppressor
module «MSM II» and the detector block is described in Section 2.8.

1 Connect column to injection valve

Remove end caps from column 73.

Screw inlet end of separating column 73 (note flow direction)

to column connection capillary 26 or to the already installed

precolumn (procedure see Section 2.7.4).

2 Rinse column

Place a beaker beneath the column outlet.

Open software window for manual system control.

If necessary, modify

Flow rate

to the value suited for the inser-

ted separating column and click on

<Send to unit>

to send this

value to the 861 Advanced Compact IC.

Switch on high-pressure pump (

IC pump

) by clicking

<On>

and rinse column with eluent for approx. 10 min.

Switch off high-pressure pump by clicking

<Off>

.

25 Mounting rail

for column holder 74

26 Column connection capillary

Connection injection valve 28 –

separating column 73

28 Injection valve

44 Inlet capillary for detector

block

PEEK capillary, fixed mounting

45 Detector block (1.733.0110)

73 Separating column

74 Column holder (6.2027.0X0)

45

44

25

28

73

25

74

26

74