2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

46

Select

Connect to workplace

in the

Control

menu in this window.

2 Open control window

Double-click the 861 icon in the system window. The control

window for manual control of the 861 Advanced Compact IC

appears, which indicates conductivity, pressure and current

system parameters.

Make sure that the set flow does not exceed the maximum

permissible flow for the particular column (see column data

sheet).

3 Start system

Make sure that the aspirating tubing 63 for the high-pressure

pump is immersed in the eluent.

Select

Startup hardware (Measure baseline)

in the

Control

menu

of the system window. The high-pressure pump, peristaltic

pump and, if present and activated under

Configuration

(see