2.8 Suppressor module «MSM II»

861 Advanced Compact IC / Instructions for Use 8.861.1033

41

module «MSM II» 46 from below through one of the openings

8 out of the inner compartment of the 861 Advanced Com-

pact IC.

Lead outlet capillary 87 to a sufficiently large waste container

and fix it in place.

Figure 18:

Connections at suppressor module

«MSM

II»

5 Inlet capillary for H

2

O

Attach inlet capillary 84 marked with "H

2

O" of suppressor mo-

dule «MSM II» 46 using a 6.2744.010 Compression fitting to

the PEEK Coupling 81 connected to the front pump tubing

79.

6 Outlet capillary for H

2

O

Pull outlet capillary 86 marked with "Waste" of the suppressor

module «MSM II» 46 from below through one of the openings

8 out of the inner compartment of the 861 Advanced Com-

pact IC.

Lead outlet capillary 86 to a sufficiently large waste container

and fix it in place.

7 Fasten capillaries to the side walls

If necessary, the two aspirating tubings 75 and 76 can be fi-

xed in the required position in the interior with the help of a

Y.107.0150 self-adhesive strap.

If necessary, the two outlet capillaries 86 and 87 can be fixed

in the required position in the interior with the help of a

Y.107.0150 self-adhesive strap.

82 Suppressor inlet capillary

for eluent

83 Suppressor outlet capil-

lary for eluent

84 Suppressor inlet capillary

for H

2

O

85 Suppressor inlet capillary

for H

2

SO

4

86 Suppressor outlet capil-

lary for H

2

O

87 Suppressor outlet capil-

lary for H

2

SO

4

1

3

2

86

84

83

82

85

87

Waste

Eluent

H

2

O

Detector

Waste

H

2

SO

4