4.3 Faults and malfunctions

861 Advanced Compact IC / Instructions for Use 8.861.1033

83

Malfunction

Cause

Rectification

Extreme peak
broadening,
splitting (double
peaks)

Dead volume at the co-
lumn ends

Check the connections
(use PEEK capillaries with
0.25 mm i.d. between in-
jection valve and detector
block)

Turn separation column (if
allowed), or replace it

No feed of re-
generation or
rinsing solution
for suppressor

Contact pressure too
low

Leak in system

Defective pump tubing

Contamination of the fil-
ter in the 6.2821.120 Fil-
ter unit PEEK

Counterpressure at sup-
pressor module «MSM
II» too high

Adjust contact pressure
(see Section 2.10.2)

Check connections and
make leakproof

Replace pump tubing (see
Section 4.2.10)

Replace the 6.2821.130
Filter (see Section 2.3.6)

Clean or replace suppres-
sor (see Section
4.2.7

4.2.9)