2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

18

2.3.2

Connection of syringe and aspirating tubing

For manual filling of the sample loops 27 mounted on the injection

valves, the 6.2816.020 Syringe and the PTFE aspirating tubing 29 al-

ready screwed to the valve are needed. These accessories are

mounted or adjusted as follows:

1 Connect syringe

Remove the plastic stopper from connection socket 3 on the

front side of the 861 Advanced Compact IC (see Figure 1).

Push 6.2816.020 Syringe (without needle) as far as it will go

into connection socket 3.

2 Lead aspirating tubing to the outside

Lead the PTFE aspirating tubing 29 connected to the injec-

tion valve 28 through one of the feedthroughs 4 to the out-

side.

2.3.3

Connection of the drain tube for the inner compartment

The 861 Advanced Compact IC has a connection at the front to which a

drain tube for discharged liquids in the inner compartment can be at-

tached. Proceed as follows:

1 Connect drain tube

Mount 6.1816.020 Silicone tubing on connection nipple on

the lower right side.

2 Lead drain tube to collecting vessel

Lead the other end of the drain tube to a suitable collecting

vessel and fix in place.

2.3.4

Connection of the drain tube for bottle rack

The 861 Advanced Compact IC has a connection at the rear to which a

drain tube for discharged liquids in the bottle rack can be attached.

Proceed as follows:

1 Connect drain tube

Mount 6.1816.020 Silicone tubing on connection nipple 9

(see Figure 2).

2 Lead drain tube to collecting vessel

Lead the other end of the drain tube to a suitable collecting

vessel and fix in place.

2.3.5

Connection of PEEK capillaries

For the connections between high-pressure pump and detector block

6.1831.010 PEEK capillaries (i.d. = 0.25 mm, o.d. =

1

/

16

"

) are used

which are connected using either 6.2744.010 PEEK compression fit-

tings (long) or 6.2744.070 PEEK compression fittings (short).

Advertising