4.2 Maintenance and servicing

861 Advanced Compact IC / Instructions for Use 8.861.1033

71

rinsed free from salt with methanol/water (1:4) to avoid crystallization

of eluent salts with the corresponding subsequent damage.
For rinsing the connections to the separating column and the suppres-

sor module are removed; the two capillaries 26 and 44 (see Figure 14

and Figure 16) are directly connected to each other using a coupling.

Rinse with methanol/water (1:4) until the conductivity drops below

10

µ

S/cm.

Before taking it back into operation, rinse the system with eluent (be-

fore the connection of the separation column).

4.2.5

Changing separating columns

It is recommended that the precolumn is also changed each time you

change the separation column).

Identical separation system
If you wish to replace an IC separating column by a column of the same

type, proceed as follows (see Figure 14 and Figure 16):

1 Remove old column

Switch off high-pressure pump and wait for pressure to drop.

Unscrew column 73 from inlet capillary 44 of the detector

block or from suppressor inlet capillary 82.

Unscrew column 73 from column connection capillary 26 or

the precolumn.

2 Connect new column to injection valve

Remove end caps from column 73.

Screw inlet end of separating column 73 (note flow direction)

to column connection capillary 26 or to the precolumn (see

Section 2.7.4/2.7.5).

3 Rinse column

Place beaker beneath the column outlet.

Switch on high-pressure pump and rinse column with eluent

for approx. 10 min, then switch off pump.

4 Connect column to detector block

Screw outlet end of separating column 73 to inlet capillary 44

or suppressor inlet capillary 82.

Changing the separation system
If you wish to replace an IC separating column by a column of a diffe-

rent type, proceed as follows (see Figure 14 and Figure 16):

1 Remove old column

Switch off high-pressure pump and wait for pressure to drop.

Unscrew column 73 from inlet capillary 44 of the detector

block or from the suppressor inlet capillary 82.