4.2 Maintenance and servicing

861 Advanced Compact IC / Instructions for Use 8.861.1033

81

Insert the new pump tubing 78 or 79 in the tubing cartridge

47 as shown in Figure 15. The orange-yellow stopper 80

must click into the corresponding holder on the left-hand side

of the tubing cartridge.

Place the tubing cartridge 47 on mounting pin 52 and press

down on the right-hand side until snap-action lever 50 clicks

into position on holding clamp 49. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

Switch on peristaltic pump.

Press contact pressure lever 48 upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.

Switch off peristaltic pump.