3 connecting the eluent bottle, Connecting the eluent bottle, Figure 12 – Metrohm 861 Advanced Compact IC User Manual

Page 36: Connection of the pulsation dampener, To protect the column, Figure 12 ), Pulsation dampener 60 i, Connection 62 of the pulsation dampener 60, Connection 61 of the pulsation dampener 60

Advertising
background image

2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

26

Figure 12: Connection of the pulsation dampener

24 Inlet capillary for injection valve

PEEK capillary 6.1831.010,

length L = 24 cm

28 Injection valve

33 Inlet capillary for injection valve

PEEK capillary 6.1831.010,

length L = 13 cm

34 Filter unit PEEK (6.2821.120)

35 Connection capillary

6.1831.010 PEEK capillary,

length L = 13 cm

37 Purge valve

60 Pulsation dampener (6.2620.150)

61 Connection to injection valve

62 Connection to purge valve

2.6.3

Connecting the eluent bottle

The eluent supply line from the storage bottle to the high-pressure

pump is connected as follows (see Figure 13):

Only

degassed (with N

2

, He or vacuum) and

microfiltered (0.45

µ

m

filter)

eluents should be used!

The 6.1608.070 Eluent bottle (2 L) supplied is

not suitable for vacu-

um degassing. Use a pressure-resistant container for this.

Care must be taken that the

eluent used is freely miscible with any

solvent remaining in the pump head (the pump head leaves the fac-

24

35

37

34

33

62

61

28

60