10 replacing the pump tubing, Replacing the pump tubing, Ction 4.2.10 ) – Metrohm 861 Advanced Compact IC User Manual

Page 90

Advertising
background image

4 Notes - Maintenance - Faults

861 Advanced Compact IC / Instructions for Use 8.861.1033

80

one of the openings of the holder.

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

Rotors which are not inserted correctly can be

destroyed

when starting operation.

5 Clean connection piece

Clean the sealing surface of new connection piece 112

(6.2832.010) with the aid of a lint-free cloth and ethanol.

6 Insert connection piece

Insert new connection piece 112 in suppressor holder 114 in

such a way that connection "1" is at the top and that the three

lugs on the connection piece fit in the corresponding open-

ings of the holder.

Screw nut 111 onto the thread of suppressor holder 114

manually (do not use tools).

7 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.

4.2.10 Replacing the pump tubing

The pump tubings used by the peristaltic pump are consumable mate-

rial with a limited lifetime and should be exchanged at regular intervals

(approx. every 4 weeks under continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in Section 2.10.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges 47 should

be raised completely by loosening the snap-action lever 50 on the right-

hand side (the set contact pressure remains unchanged).
As the pump is always operated on the same side the 6.1826.110

Pump tubings supplied can be used on both sides. To exchange a

pump tubing proceed as follows:

1 Remove old pump tubing

Press contact pressure lever 48 on the tubing cartridge 47

down as far as it will go.

Release tubing cartridge 47 from holding clamp 49 by press-

ing down snap-action lever 50 and remove from mounting pin

52 (see Figure 16).

Remove old pump tubing 78 or 79.

2 Insert new pump tubing

Advertising