2.7 Precolumns and separating columns

861 Advanced Compact IC / Instructions for Use 8.861.1033

31

1 Connect precolumn

Remove end caps from the precolumn.

Fit compression fitting to the connection capillary 26 moun-

ted to the injection valve (see Section 2.3.5).

Screw column connection capillary 26 to precolumn.

Cut off a small piece, approx. 5 cm, from the 6.1831.010

PEEK capillary and fit compression fittings to both ends (see

Section 2.3.5).

Mount the prepared capillary to the outlet of the precolumn.

When you install the column always ensure that this is inserted cor-

rectly in accordance with the flow direction (if existent) shown.

2 Rinse the precolumn

Place a beaker beneath outlet capillary.

Switch on IC Pump in «IC Net» and rinse precolumn for

approx. 10 min with eluent.

Switch off IC Pump.

2.7.2 General

information

on separating columns

The precolumn that is suitable for your separating column can be

found in the

Metrohm IC Column Catalog, the product information

about separating columns that can be found under

http://www.met-

rohm.com

, ion chromatography products, or let your agency advise

you directly.

New IC columns are normally supplied filled with liquid and sealed at

both ends. Before inclusion in the system you must make sure that

this solution is fully miscible with the eluent to be used (observe the in-

formation provided by the manufacturer).

The separating columns and precolumns currently available from
Metrohm can be found in the Metrohm IC Column Catalog, or on the

Internet under

http://www.metrohm.com

in the ion chromatography pro-

duct section. Each column is supplied with a test chromatogram and a
data sheet. More detailed information about special IC applications can
be found in the relevant "Application Bulletins" or "Application No-

tes"; these are available on the Internet under

http://www.metrohm.com

in the applications sector, or can be requested free of charge from your
local Metrohm agency.

When you install the column always ensure that this is inserted cor-

rectly in accordance with the flow direction shown.

2.7.3

Selecting the sample loop

Selection of the sample loop depends on the separating column used.

Normally the following sample loops are used: