4 Notes - Maintenance - Faults

861 Advanced Compact IC / Instructions for Use 8.861.1033

78

the connection piece 112 (order number 6.2832.010).

4 Clean suppressor rotor

Clean the sealing surface of suppressor rotor 113 using a lint-

free cloth and ethanol.

5 Insert suppressor rotor

Insert suppressor rotor 113 in suppressor holder 114 in such

a way that the tubing connections at the rear of the rotor fit in

the corresponding openings inside the rotor, and that one of

the three holes in the rotor can be seen from below in one of

the openings of the holder.

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

Rotors which are not inserted correctly can be

destroyed

when starting operation.

6 Clean connection piece

Clean the sealing surface of connection piece 112 using a lint-

free cloth and ethanol.

7 Insert connection piece

Insert connection piece 112 in suppressor holder 114 in such

a way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

Screw nut 111 onto the thread of suppressor holder 114

manually (do not use tools).

Figure 22: Assembling the suppressor

111

112

113

114

115