3 injection valve, 4 high-pressure pump, Injection valve – Metrohm 861 Advanced Compact IC User Manual

Page 96: High-pressure pump

Advertising
background image

5 Appendix

861 Advanced Compact IC / Instructions for Use 8.861.1033

86

Max. temperature deviation

±

2.5°C

Heating time

30 min

Temperature stability

0.01°C at constant ambient temperature

Connection for detector

block

Dsub 15 pin (female)

5.1.3 Injection

valve

Actuator switching duration 100

150 ms

Pressure resistance

35 MPa (350 bar)

5.1.4 High-pressure

pump

Type

Serial dual piston pump with two valves

Pump capacity

Flow range

0.20...2.5 mL/min

Maximum error

<

±

2 % of set value

Flow constancy

< 0.5 % of set value

Reproducibility of eluent

flow

typ. better than

±

0.1 %

Pressure measurement

Pressure range

at flow 0.2 ... 1.5 mL/min:

0 ... 35.0 MPa (0 ... 350 bar)
at flow 1.5 ... 2.5 mL/min:

0 ... 25.0 MPa (0 ... 250 bar)

Residual pulsation

< 10 % (at 1 mL/min water and 10 MPa pressure,

without pulsation dampener)

Measurement principle

Piezoresistive measurement

Response time: 3 ms

Measurement volume: approx. 50

µ

L

Maximum error

±

3 % of set value

Resolution

0.1 MPa (conductivity measurements)

0.01 MPa (pressure measurements)

Sampling rate

1 measurement/piston stroke (pump running)

1 measurement/s (pump not running)

10 measurements/s (pressure measurements)

Safety shutdown

Function

Automatic shutdown when upper and lower pres-

sure limits violated

Maximum pressure limit

adjustable between 0.1...35.0 MPa (1...350 bar)

Response time: 1 pump cycle

Advertising