4 Notes - Maintenance - Faults

861 Advanced Compact IC / Instructions for Use 8.861.1033

70

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the inte-

rior of the instruments the mains plug of the 861 Advanced Compact

IC must be immediately disconnected to prevent extensive damage to

the instrument electronics. Inform Metrohm service if your instru-

ment(s) have been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in Section 1.4.1.

Maintenance by Metrohm service
Maintenance of the 861 Advanced Compact IC is best done as part of

an annual service performed by specialists from the Metrohm compa-

ny. If work is frequently performed with caustic and corrosive chemicals,

it may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.2.2 Passivation

Passivation of the complete IC System (without column) by rinsing with

20...50 mL 0.2 mol/L HNO

3

is only required when unusual alterations to

the measuring properties of the cell are observed. For passivation the

separating column 73 is removed from the 861 Advanced Compact IC.

The two in- and outlet capillaries are directly connected to each other

using the Coupling (6.2620.060).

4.2.3 Recycling

To keep the eluent consumption between injections to a minimum when

the system is at rest (e.g. overnight), the so-called recycling procedure

can be used. In recycling the eluent exiting the outlet capillary of the de-

tector block is led back directly to the eluent supply vessel 67. The IC

system is thus quickly ready for new injections without a long condition-

ing period.
Instead of threaded stopper 65 the 4.420.0311 Tubing nipple (M6),

which belongs to the accessories of the 6.1602.160 Bottle attachment,

can be used together with the second E.301.0021 O-ring (which is also

included) for returning the eluent (see Section 2.6.3).

The recycling procedure must

not be used

in operation with the suppressor module «MSM II»

with alkaline eluents

with the 6.1010.000 IC Cation column METROSEP Cation 1-2.

4.2.4 Shutdown

If the 861 Advanced Compact IC is shut down for a considerable length

of time, the entire IC system (without column and suppressor) must be