2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

28

861 Advanced Compact IC, cut off to the required length and

push at least 5 mm of it onto aspirating capillary 38 (see

Figure 3 and Figure 4 and Figure 5) of the high-pressure

pump (it may be necessary to use emery paper).

If necessary, fix aspirating tubing 63 in the required position

in the interior with the aid of the Y.107.0150 self-adhesive

strap.


Figure 13: Connection of elu-

ent bottle

2.6.4

Deaerating the pump and rinsing the pulsation dampener

The first time that it is started up the high-pressure pump must be de-
aerated. Proceed as follows:

1 Prepare for deaeration

Open the rotary knob on purge valve 37 by approx. ½ turn in

the counterclockwise direction (see Figure 3 and Figure 4 and

Figure 5).

Remove the plastic stopper from connection 2 on the front

panel of 861 Advanced Compact IC (see Figure 1).

Push 6.2816.020 Syringe (without needle) into connection 2

until the stop is reached.

2 Open and connect system

Start the «IC Net» PC program, if it has not been already

been started (see Section 2.5.3).

Select

File / Open / System

in the main window. Select the sys-

tem created in Section 2.5.4, and click on

<Open>

.

63 Aspirating tubing (6.1834.010)

64 Tubing nipple (4.420.4300; M8)

with E.301.0021 O-ring

65 Threaded stopper (6.1446.040; M6)

66 Bottle attachment (6.1602.105)

67 Eluent bottle (6.1608.070)

68 Aspirating filter (6.2821.090)

69 CO

2

absorber

70 Cotton wool

71 SGJ clip (6.2023.020)

72 Absorber tube (6.1609.000)

68

72

71

70

69

63
64

65

66

67