2 Installation
861 Advanced Compact IC / Instructions for Use 8.861.1033
52
selected at the 813 (see 813 Instructions for Use, Section
3.4).
The two Valve commands switch the injection valve of the
861 to the positions Fill and Inject.
For more detailed information about the control please refer
to Section 3.3.1 or the «IC Net» Software Instructions for Use.
In order for the 861 – 813 connection to function properly Re-
mote line 3 at the 861 Advanced Compact IC must be set to 0
each time that a determination is started (
System startup values
:
Set
Remote line 3 = 0
).