Metrohm 861 Advanced Compact IC User Manual

Page 73

Advertising
background image

3.3 Settings

861 Advanced Compact IC / Instructions for Use 8.861.1033

63

Thermostat

Operating temperature of the conductivity

cell.
Selection:

25

,

30

,

35

,

40

,

45 °C

,

off

Thermostatting functions only if the ambient tem-

perature is at least 5°C lower than the operating

temperature. It normally takes 30

60 min after

power on until a temperature stability of

±

0.01 °C is

attained.

Flow correction

Factor for correction of the difference be-

tween displayed and actual flow rate of
the high-pressure pump.

Range:

0.9

...

1.09

The correction fac

tor is determined as

follows by measuring the actual flow rate
using a graduated cylinder:

Flow correction

=

Displayed flow rate
Measured flow rate

Cell constant

Cell constant of the conductivity cell for

correct display of the absolute conductiv-
ity. If you connect a new detector block
enter the cell constant printed on the de-
tector block into this field.

Range:

0.1

...

1000 /cm

For a precise determination of the cell
constant, pump a calibration solution of
known conductivity through the IC sys-

tem, observe the displayed conductivity
and change the cell constant until the
correct conductivity value is displayed.

<Break times>

Possibility for changing the break times

for injection valve

Valve

and suppressor

module

Suppressor

.