Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 109

Advertising
background image

▪▪▪▪▪▪▪

107

When Vision first communicates with the
instrument (or after disconnection and re-
connection) this screen is shown.

Vision must establish the configuration and
have it verified by the user. This assures
that test data is sent to the correct location
in the Diagnostic Database.

If the Sample Module Serial Number field is
empty, locate the sample module serial
number on the side of the module. Record
the serial number and enter it in this field.

Click “OK to accept the instrument
identification.

Vision briefly displays a dialog box which indicates that the instrument reference is being placed into
position. In reflectance, this is automatic. No user action is required.

The status bar indicates scan progress.

Vision requests that the Certified 80%
Reflectance Reference Standard be placed
in the sampling area.

Place the standard as shown, with the
identifying label parallel to the long axis of
the tray.

Close the cover, and click on “OK”.