Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 118

Advertising
background image

116

▪▪▪▪▪▪▪

Information about the reflectance wavelength standard:

The wavelength standard used is directly traceable to NIST SRM-1920a, through direct comparison
on the Metrohm master reflectance instrument, and in chemical formulation. In addition to the
prescribed formulation, one additional ingredient is added, in a small amount, to provide peaks
beyond those normally found in SRM-1920a. This material has very sharp bands, which are found to
be stable and repeatable.

Spectra of each are as described:

The darker spectrum, which has no discernible peaks beyond about 2150nm, is SRM-1920a.

The FOSS WSR Wavelength Standard is the lighter spectrum, and has clear peaks visible at above
2200nm. These additional peaks are used to set the wavelength scale of the instrument to aid in
instrument matching. This is one important step in method transfer.

The Foss WSR Wavelength Standard exhibits slightly different absorbance and baseline levels, due to
the reflectivity of the added ingredient. The peaks, however, are in the same wavelength positions,
and are similar in shape to SRM-1920a.

The XDS instrument is set to NIST nominals during Instrument Calibration, since these are the best
known information for NIR peak wavelengths. Peaks for the additional ingredient are set to peak
nominals determined by measurement on several different types of research instrumentation.

In Wavelength Certification, the NIST-stated uncertainty of 1.0nm is applied. Tighter tolerances are
not appropriate, unless NIST revises the stated uncertainty of SRM-1920a at some point in the future.

NOTE: The response of the WSR Wavelength Standard may vary slightly with temperature. This is
typically in the range a few hundredths of a nanometer for small temperature variations. While this
effect is small, it may cause some variation when running Wavelength Certification.

We suggest that the WSR standard be stored in the standards box, rather than inside the instrument.
When running a test with the WSR, place it in the instrument as directed, then take it out and store it
in the box, to keep the temperature as consistent as possible. Temperature inside the instrument may
be as much as ten degrees (F) higher, and may cause slight wavelength drift, enough to cause slight
wavelength errors when measuring wavelength response.

Advertising