instrument operation. Please proceed to the next step.

In tablet transmission, Vision stores yet another Reference Standardization, for the transmission
reference. This is addressed in the section covering Tablet Transmission.

6.2

Use of Vial Dialog Selection Menu

The MasterLab Analyzer is designed for automated scanning of multiple tablets or vials. Vision
Software supports this feature, and permits the user to set up the system for each sample to be
collected. While the instrument is scanning, the user can prepare the next set of samples, or perform
other necessary lab work.

This section uses a vial tray, in reflectance, to explain menu commands. When using a DCM in tablet
transmission, the default screen is slightly different, and the positions are named “Tablet 1”, and so
on.

Please see page 44 for specifics about tablet analysis in transmission.

To set up the MasterLab Analyzer for automated sampling, follow these instructions:

From the menu bar, go to Acquire, Sample.

Vision prompts for the Product ID using this
dialog box. Enter the correct product name in the
Product ID field.

NOTE: If multiple products exist, click the arrow
to see all products, then highlight and select the
desired Product ID.

The Sample ID Prefix is not entered in this box –
that is done in the next screen.

Click “OK” when finished.

Vision displays this screen, which corresponds to the tablet (or vial) tray. Note that the Product ID is
entered -- this came from the previous dialog box.