Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 94

Advertising
background image

92

▪▪▪▪▪▪▪

Double-click the lower right quadrant to see the full report.

The first tab shows the Wavelength position, as found. Only one is used with this detector, which is
very linear and predictable in response. The peak position is compared to the nominal peak position
as identified on an instrument calibrated using NIST SRM-2035. The difference from nominal, and the
repeatability of position are calculated. The acceptance specification is +/-1.0nm.