2 creation of a custom vial size, Creation of a custom vial size – Metrohm NIRS XDS MasterLab Analyzer User Manual
Page 33
▪▪▪▪▪▪▪
31
5.2
Creation of a Custom Vial Size
Most users will work with one of the standard vial sizes (or tablet layouts) offered. However, for
setup of custom sizes, this information is provided. To create a custom vial diameter, follow these
steps:
From the menu bar, select Configure, Tray.
At the Select Tray Dialog box, note that four
sizes (vial diameters) are already loaded in
Vision.
NOTE: The spot size (set in the Data Collection
Method) should be no larger than 80% of the
vial diameter. For each default vial size, set spot
size as follows:
15mm vial:
12mm spot size
19mm vial:
15mm spot size
25mm vial:
17.5mm spot size
29mm vial:
17.5mm spot size
If your vial diameter is not shown, ignore the
default sizes. Click “New”.
This brings up the Tray Configuration Dialog
box shown below.