Vision requests the “Standard File” for the Certified
Wavelength Standard.

Place the mini-CD into the CD drive, and select the
standard file. Click “Open”.

The file may be loaded into the C:\Vision directory, and
recalled from that location.

Vision acquires another instrument reference.

Vision prompts the user to insert the Tablet
Transmission Standard Tray into the instrument.

Insert the tablet standards tray into the sample carriage
when prompted. (If this was not already done.)

Click “OK” to continue.

Vision acquires a spectrum of the Wavelength Standard
and performs several operations.

This test takes about 45 seconds.